Laser & Optoelectronics Progress, Volume. 54, Issue 12, 122501(2017)

Fast Radiometric Calibration Method for Long Wave Infrared Detectors

Nan Tongling1,2, Shen Honghai1、*, Yang Mingyu1, Chen Cheng1, and Zhu Dantong1,2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Using high-accuracy extend blackbody as the standard radiation source, we set up the radiometric calibration system and perform a radiometric calibration experiment for long wave infrared camera. Based on the practical response data, the relationship of response characteristics with incident radiation and integration time is analyzed, and a mathematical model of radiometric calibration is establish. Thereout, the fast radiation calibration method, which is choose of three pictures under two temperatures at two typical integral time, is proposed. The unknown parameters in the calibration model can be obtained by data processing. Finally, using 398 sets of available data from 420 sets of measured data, we verify the reliability of the simplified radiation calibration model. The results show that the relative error between the calculated and measured values is less than 1%, and the fitting degree is greater than 0.999. Experiments show that the calibration method can ensure the accuracy of rapid radiation calibration under different integral time.

    Tools

    Get Citation

    Copy Citation Text

    Nan Tongling, Shen Honghai, Yang Mingyu, Chen Cheng, Zhu Dantong. Fast Radiometric Calibration Method for Long Wave Infrared Detectors[J]. Laser & Optoelectronics Progress, 2017, 54(12): 122501

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: OPTOELECTRONICS

    Received: Jul. 10, 2017

    Accepted: --

    Published Online: Dec. 11, 2017

    The Author Email: Honghai Shen (shenhh@ciomp.ac.cn)

    DOI:10.3788/lop54.122501

    Topics