Infrared and Laser Engineering, Volume. 44, Issue 10, 2959(2015)

Standard scale calibration methods for δ13CO2 value measurements based on Fourier Transform Infrared spectroscopy

Li Xiangxian1、*, Luo Guishan2, Xu Liang1, Gao Minguang1, Tong Jingjing1, Wei Xiuli1, and Liu Na1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Calibration to standard reference gas scale is necessary to improve the accuracy of the measurements,and ensure the compatibility of measurement values between different instruments for δ13CO2 value based on Fourier Transform Infrared(FTIR) spectroscopy. Firstly, the calibration methods were put forward, the absolute calibration and empirical calibration methods were studied in detail. On the basis of these researches, the δ13CO2 values were calibrated to standard reference gas scale. The calibration results show that measurements accuracy has been improved greatly, compared with measurement results of no calibration. The accuracy for absolute calibration and empirical calibration methods are improved 5.5 and 6.4 times respectively. The study is benefit to improve the accuracy of FTIR measurements and the compatibility between different instruments.

    Tools

    Get Citation

    Copy Citation Text

    Li Xiangxian, Luo Guishan, Xu Liang, Gao Minguang, Tong Jingjing, Wei Xiuli, Liu Na. Standard scale calibration methods for δ13CO2 value measurements based on Fourier Transform Infrared spectroscopy[J]. Infrared and Laser Engineering, 2015, 44(10): 2959

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: 光电测量

    Received: Feb. 10, 2015

    Accepted: Mar. 15, 2015

    Published Online: Jan. 26, 2016

    The Author Email: Xiangxian Li (lixx@aiofm.ac.cn)

    DOI:

    Topics