Laser & Optoelectronics Progress, Volume. 57, Issue 18, 181701(2020)

Dual-Field Capsule Endoscope Design Based on Zemax Software

Jinxuan Huang*, Yanfang He, and Feng Lin**
Author Affiliations
  • 2Key Laboratory of Medical Optoelectronic Science and Technology, Ministry of Education, College of Optoelectronics and Information Engineering, Fujian Normal University, Fuzhou, Fujian 350007, China
  • show less

    The capsule endoscope currently on the market only images the front field of view. Furthermore, the folds and loops of intestinal tissue in the human body cause a large number of blind spots in the rear field of view. In this paper, we report the design of a capsule objective lens that simultaneously images the front (0°-80°) and the side rear (50°-80°) fields of view. The objective lens is 15.6 mm in length, and the maximum radial diameter is 8 mm. An OH02A10 imaging chip ensures a high-resolution image (maximum cut-off frequency >0.5 at 89 lp/mm) and provides good image quality.

    Tools

    Get Citation

    Copy Citation Text

    Jinxuan Huang, Yanfang He, Feng Lin. Dual-Field Capsule Endoscope Design Based on Zemax Software[J]. Laser & Optoelectronics Progress, 2020, 57(18): 181701

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Medical Optics and Biotechnology

    Received: Oct. 28, 2019

    Accepted: Dec. 31, 2019

    Published Online: Sep. 2, 2020

    The Author Email: Huang Jinxuan (huangjx4267@qq.com), Lin Feng (lfeng127@163.com)

    DOI:10.3788/LOP57.181701