Chinese Optics Letters, Volume. 5, Issue 8, 457(2007)

Method for changing brightness temperature into true temperature based on twice recognition method

Yang Song, Xiaogang Sun, and Hong Tang*
Author Affiliations
  • Department of Automation Measurement and Control, Harbin Institute of Technology, Harbin 150001
  • show less

    The channel output of a multi-wavelength pyrometer is the brightness temperature rather than the true temperature. Twice recognition method is put forward to change the brightness temperatures of a multi-wavelength pyrometer into the true temperatures of targets. Using the data offered by Dr. F. Righini, the experimental results show that the difference between the calculated true temperature based on twice recognition method and the real true temperature is within +-20 K. The method presented in this paper is feasible and effective for the true temperature measurement of targets.

    Tools

    Get Citation

    Copy Citation Text

    Yang Song, Xiaogang Sun, Hong Tang. Method for changing brightness temperature into true temperature based on twice recognition method[J]. Chinese Optics Letters, 2007, 5(8): 457

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Nov. 9, 2006

    Accepted: --

    Published Online: Aug. 8, 2007

    The Author Email: Hong Tang (tangbenben@126.com)

    DOI:

    Topics