Optics and Precision Engineering, Volume. 20, Issue 10, 2192(2012)

Design and experiment of intermittent working image capsule endoscopy

LIU Gang*... YAN Guo-zheng, LIU Hua and ZHU Bing-quan |Show fewer author(s)
Author Affiliations
  • [in Chinese]
  • show less

    As the image capsule endoscopy is limited by the energy and it has short working time, this paper designs a miniature intermittent working image capsule endoscopy with low power consuming according to the speed of image capsule's peristalsis in human intestine. The Hall switch was designed as the main power supply switch instead of reed switch to further reduce the volume of the image capsule. A battery power supplies module, an analog switch circuit, and a software were developed to realize the intermittent power supplies of wireless transmitting and Light Emitting Diode(LED) circuits and the sleep mode of an image sensor. The clear images were obtained successfully, and the tests and animal experiments were performed to prove the work time and feasibility of the whole system. Obtained results show that the total size of the system is only Φ12 mm×25 mm, and the work time is more than 9 h in one second intermittent working mode when the modules are connected.

    Tools

    Get Citation

    Copy Citation Text

    LIU Gang, YAN Guo-zheng, LIU Hua, ZHU Bing-quan. Design and experiment of intermittent working image capsule endoscopy[J]. Optics and Precision Engineering, 2012, 20(10): 2192

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Jun. 9, 2012

    Accepted: --

    Published Online: Nov. 1, 2012

    The Author Email: Gang LIU (liugang531557127@163.com)

    DOI:10.3788/ope.20122010.2192

    Topics