Laser & Optoelectronics Progress, Volume. 52, Issue 7, 73006(2015)

Precision Improvement of Laser-Induced Breakdown Spectroscopy Using Relevant Lines as Internal Standard

Liu Li*
Author Affiliations
  • [in Chinese]
  • show less

    To improve the analytical precision of laser-induced breakdown spectroscopy (LIBS), a modified internal standard approach is studied based on the correlation of analytical lines and reference lines. The approach is investigated in detail by principle analysis and numerical calculation firstly, then experimental verification is carried out using the analytical lines of Co I 350.23 nm and Cr I 427.48 nm from the micro-alloy steel standard samples. The results show that, to improve the analytical precision of LIBS, it must meet the necessary conditions that the correlation coefficient of analytical line and reference line γ >0.5, the noise proportional factor b>0.5, and γb > 0.5 . The precision enhancement factor R increases with the increasing of γ , but increases first then decrease with b. When the fluctuation of analytical lines is greater than that of reference lines, the degree of precision improvement decrease gradually.

    Tools

    Get Citation

    Copy Citation Text

    Liu Li. Precision Improvement of Laser-Induced Breakdown Spectroscopy Using Relevant Lines as Internal Standard[J]. Laser & Optoelectronics Progress, 2015, 52(7): 73006

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Mar. 20, 2015

    Accepted: --

    Published Online: May. 29, 2015

    The Author Email: Li Liu (ll4246@126.com)

    DOI:10.3788/lop52.073006