Chinese Journal of Quantum Electronics, Volume. 26, Issue 5, 624(2009)

Absolute infrared flux responsivity calibration of trap detectors based on cryogenic radiometer

Yun-jun LU1、*, Xiao-bing ZHENG1, Jian-jun LI1, Wei ZHANG1, Ping XIE1, Yao NAN2, and Peng SANG2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    At the wavelength of 1064 nm, the absolute optical power responsivities of two Si trap detectors were measured against the cryogenic radiometer. The calibration method and experiment equipment were discussed, as well as analysis of the measurement data. A new quasi in-situ measurement scheme of Brewster windows transmittance was proposed and transmittance higher than 0.999 was obtained. Results show that the combined uncertainties of the responsivity are1.457×10ˉ4 and 1.458×10ˉ4, and the measurement repeatabilities of the two trap detectors were within 0.014% and 0.008%, respectively. The uncertainties and possible error sources during the process of measurement were analyzed and evaluated. The temperature stability of optical power responsivity of 0# trap detector was studied at the wavelength of 1064 nm.

    Tools

    Get Citation

    Copy Citation Text

    LU Yun-jun, ZHENG Xiao-bing, LI Jian-jun, ZHANG Wei, XIE Ping, NAN Yao, SANG Peng. Absolute infrared flux responsivity calibration of trap detectors based on cryogenic radiometer[J]. Chinese Journal of Quantum Electronics, 2009, 26(5): 624

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Oct. 10, 2008

    Accepted: --

    Published Online: May. 24, 2010

    The Author Email: Yun-jun LU (yjlu@aiofm.ac.cn)