Research on High-Accuracy in-Flight Spectral Calibration of the Solar Backscattered Ultraviolet Spectroradiometer
A new high-accuracy spectral calibration is applied to monitor and correct spectral shift of solar backscattered ultraviolet spectroradiometer (SBUS) in launch and flight. A high-resolution reference spectrum is used to convolve with instrument′s slit function. The convolved results is matched with corrected spectrum measured by SBUS, and evaluation function is founded to calculate the value of the spectral shift that minimizes the difference between reference and measured spectrum. For SBUS, the optimal spectral shift is 0.10 nm. Finally, the result is validated by standard spectrum of on-board mercury lamp and the maximal error is 0.01 nm, which meets instrument′s requirements. The experimental results verity feasibility and precision of the high-accuracy spectral calibration.
Get Citation
Copy Citation Text
Li Zhanfeng, Wang Shurong, Huang Yu, Yu Xiangyang. Research on High-Accuracy in-Flight Spectral Calibration of the Solar Backscattered Ultraviolet Spectroradiometer[J]. Acta Optica Sinica, 2013, 33(2): 228002
Category: Remote Sensing and Sensors
Received: Jun. 28, 2012
Accepted: --
Published Online: Nov. 21, 2012
The Author Email: Zhanfeng Li (lizhanfeng115@163.com)