    Creatinine (Cr) is a biochemical waste molecule generated from muscle metabolism and primarily cleared from the bloodstream by the kidneys. If kidney function declines, Cr levels in the blood tend to increase. Therefore, Cr serves as an indicator of kidney function. In this work, we present a simple method for the rapid screening for impaired renal function based on the subject's Cr concentration. In our setup, broadband white light is delivered to a finger clamp through a fiberoptic cable to illuminate the patient's finger. The light is transmitted through the finger and collected by a second optical fiber coupled to a visible–near-infrared (VisNIR) spectrometer which covers the spectral range from 400 nm to 1100 nm. During the calibration process, the transmitted spectra acquired from 60 patients were measured. An average was calculated using the peak level of the transmitted, diffused intensity at three different wavelengths to create a “Cr intensity index". Patients were divided into five groups according to their Cr concentration levels, ranging from 1 mg/dL to 13 mg/dL. Our observations indicated that each group featured a unique spectral fingerprint. Next, we tested the index on 20 patients not included in the calibration procedure (unknown samples). We were able to classify patients into groups according to their Cr level with moderate prediction accuracy (R2 = 0.55) and mean screening error of up to 16%. Future efforts will evaluate the accuracy of this approach with larger patient populations representing a broad range of Cr concentration. Still, this preliminary work is an essential step toward developing this useful noninvasive Cr screening platform using NIR light spectroscopy.

    Tools

    Get Citation

    Copy Citation Text

    Zvi Hai Barnea (Burbea), Shmuel Zimlichman, David Abookasis. Diffuse transmitted spectroscopy in conjunction with spectral peak averaging as a potential tool for noninvasive creatinine screening[J]. Journal of Innovative Optical Health Sciences, 2020, 13(6): 2050028

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: May. 10, 2020

    Accepted: Aug. 30, 2020

    Published Online: Dec. 25, 2020

    The Author Email:

    DOI:10.1142/s1793545820500285

    Topics