Chinese Journal of Quantum Electronics, Volume. 17, Issue 2, 97(2000)

Resonance Frequency Shifts due to Radiation Fields in the Laser-cooled Cesium Fountain Frequency Standard and their Effects on the Accuracy of the Standard

[in Chinese]1 and [in Chinese]2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Resonance frequency shifts due to radiation fields in the laser-cooled cesium fountain frequency standard, such as the frequency shifts due to microwave spectral impurity, Bloch-Siegert effect, neighboring π transitions, Stark effect, blackbody radiation and Majorana transitions, are investigated, with formulas for evaluating the shifts presented. The magnitudes of the shifts and their uncertainties are evaluated and their effects on the accuracy of the standard are analyzed. In addition, approaches for reducing the shifts and uncertainties are proposed in order to improve the accuracy of the standard. The results obtained here might be of great help for designing the laser-cooled cesium fountain frequency standard and evaluating the accuracy of the standard.

    Tools

    Get Citation

    Copy Citation Text

    [in Chinese], [in Chinese]. Resonance Frequency Shifts due to Radiation Fields in the Laser-cooled Cesium Fountain Frequency Standard and their Effects on the Accuracy of the Standard[J]. Chinese Journal of Quantum Electronics, 2000, 17(2): 97

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Special Issue:

    Received: Mar. 29, 1999

    Accepted: --

    Published Online: May. 15, 2006

    The Author Email:

    DOI:

    Topics