Acta Photonica Sinica, Volume. 35, Issue 12, 1832(2006)
Design of a Femoto-second Streak Tube
A femoto-streak image tube has been designed,in which a static electrode was used as the focus lens. The distribution of the initial kinetic energy,the initial azimuth angle,elevation angle and position of the photoelectrons were sampled with Monte Carlo method. The spatial electric field strength was calculated with Finite Difference method,and the trajectories of 3000 photoelectrons were calculated with fourth Runge-Kutta method. After analysis the spatial and temporal distribution of the electrons on the optimum plane,the basic character parameters such as the spatial transfer function and temporal resolution et al of the image tube were provided,the total temporal resolution of the image tube was expected to reach to 290.4 fs.
Get Citation
Copy Citation Text
Tian Jinshou, Bai Yonglin, Liu Baiyu, Ouyuan Xian, Bai Xiaohong, Yang Wenzheng, Wang Chen. Design of a Femoto-second Streak Tube[J]. Acta Photonica Sinica, 2006, 35(12): 1832
Category: Optoelectronics
Received: Sep. 1, 2005
Accepted: --
Published Online: Jun. 3, 2010
The Author Email: Jinshou Tian (tianjs@hotmail.com)
CSTR:32186.14.