Spectroscopy and Spectral Analysis, Volume. 33, Issue 5, 1180(2013)

Influence of Delay Time on Laser-Induced Breakdown Spectroscopy of Limestone

WANG Yang-en
Author Affiliations
  • [in Chinese]
  • show less

    In order to study the influence of delay time on laser-induced breakdown spectroscopy(LIBS), the relationship between the intensities of the spectral line and the delay time was discussed in theory, and the LIBS instrument manufactured by Applied Spectra Inc was used for testing the limestone certified reference sample GBW07716. The characteristic spectral line was analyzed with MnⅠ357.788 nm, WⅡ364.141 nm, WⅠ407.436 nm, WⅠ426.938 nm and CoⅠ387.312nm.It was demonstrated that the intensities of the spectral line decrease exponentially with the delay time, and the decrease speed is not the same for different spectral line. The intensities of the ion spectral line are stronger in an early stage, but decrease faster with the increase in delay time than the atomic spectral line. The experimental data fit with exponential law and the fitting R value is about 0.99.

    Tools

    Get Citation

    Copy Citation Text

    WANG Yang-en. Influence of Delay Time on Laser-Induced Breakdown Spectroscopy of Limestone[J]. Spectroscopy and Spectral Analysis, 2013, 33(5): 1180

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Oct. 8, 2012

    Accepted: --

    Published Online: May. 21, 2013

    The Author Email:

    DOI:10.3964/j.issn.1000-0593(2013)05-1180-05