Journal of Atmospheric and Environmental Optics, Volume. 12, Issue 6, 456(2017)
Measurement, Correction and Precision Verification of Out-of-Band Response for Multispectral Remote Sensor
Out-of-band response (out-of-band light leakage) is a significant factor affecting the radiometric calibration uncertainty of multispectral remote sensor (MsRS). For the MsRs out-of-band response measurement, correction and precision verification, the 443 nm band in multispectral imager using the functions of a novel spectrum-tunable integrating spheres reference light source (STIS) (i.e. discrete waveband and continuous spectrum output functions) was taken for example in batch experiments. After carrying out the out-of-band response measurement, correction and precision verification of multispectral imager, the weighted radiant luminance was amended based on the out-of-band response correction model, and compared with the measuring reference of spectral radiance meter. The results show that the difference between in-band weighted radiant luminance and measuring reference before and after the out-of-band response correction decreases from 8.8% to 2.9%, which confirms the application feasibility of STIS on the MsRS out-of-band response correction.
Get Citation
Copy Citation Text
KANG Qing, YUAN Yinlin, LI Jianjun, ZHENG Xiaobing, WU Haoyu, ZHAI Wenchao, QIAN Honghu, LUO Donggen, YANG Weifeng, HONG Jin. Measurement, Correction and Precision Verification of Out-of-Band Response for Multispectral Remote Sensor[J]. Journal of Atmospheric and Environmental Optics, 2017, 12(6): 456
Category:
Received: Aug. 23, 2016
Accepted: --
Published Online: Dec. 4, 2017
The Author Email: Yinlin YUAN (ylyuan@aiofm.ac.cn)