Acta Optica Sinica, Volume. 38, Issue 9, 0912004(2018)

Splicing Detection Technique of Homogeneity of Large-Aperture Laser Neodymium Glass

Yunbo Bai1、*, You Zhou1, Shijie Liu1、*, Jianda Shao1, and Longbo Xu2
Author Affiliations
  • 1 Key Laboratory of Materials for High Power Laser, Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Shanghai 201800, China
  • 2 School of Electronic and Optical Engineering, Nanjing University of Science and Technology, Nanjing, Jiangsu 210094, China
  • show less

    A measurement technique of one-dimensional (1D) sub-aperture splicing based on an interferometer with the aperture of 600 nm is proposed. With the analysis of pixel offset error simulation results, a 1D large air-float precision scanning splicing platform is designed and developed, the splicing detection of full aperture optical homogeneity of a large aperture laser glass with a diagonal close to 1 m is realized, and the accuracy and the repeatability of this technique are tested. The results show that the splicing results are smooth, the relative error between the homogeneous splicing and direct measurement results is superior to 2×10 -7, the repeatability of full aperture homogeneity is superior to 4×10 -8, and the largest difference between the measurement homogeneity results of single-aperture and the same region after splicing is 2×10 -7.

    Tools

    Get Citation

    Copy Citation Text

    Yunbo Bai, You Zhou, Shijie Liu, Jianda Shao, Longbo Xu. Splicing Detection Technique of Homogeneity of Large-Aperture Laser Neodymium Glass[J]. Acta Optica Sinica, 2018, 38(9): 0912004

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Mar. 2, 2018

    Accepted: Apr. 23, 2018

    Published Online: May. 9, 2019

    The Author Email:

    DOI:10.3788/AOS201838.0912004

    Topics