Acta Photonica Sinica, Volume. 45, Issue 7, 70712001(2016)

Phase-shifting Calibration Method of Wavelength-shifting Interference Measurement

TIAN Ai-ling1、*, MA Di1, WANG Da-sen2, LIU Jian1, LIU Bing-cai1, and ZHU Xue-liang1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Based on the wavelength tuning phase-shifting technology in Fizeau interferometer, a phase-shifting calibration method with step by step was presented. First, the n groups interferograms with t step were generatd , the step is π/2. Then, the error function for each step was built through the origin interferogram multiplied by n group of each step separately and making Fourier transform. Finally, the precision calibration was realized by calculating the extreme point of error function curve. The experiment results show that, the calibrating precision is better than the traditional ones and the calibration accuracy is up to 2 or 3 orders to compare with the traditional method. This new method not only avoids the random error, but also eliminates the accumulated error completely.

    Tools

    Get Citation

    Copy Citation Text

    TIAN Ai-ling, MA Di, WANG Da-sen, LIU Jian, LIU Bing-cai, ZHU Xue-liang. Phase-shifting Calibration Method of Wavelength-shifting Interference Measurement[J]. Acta Photonica Sinica, 2016, 45(7): 70712001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Jan. 14, 2016

    Accepted: --

    Published Online: Aug. 18, 2016

    The Author Email: Ai-ling TIAN (tian21964@sohu.com)

    DOI:10.3788/gzxb20164507.0712001

    Topics