Acta Optica Sinica, Volume. 31, Issue s1, 100501(2011)

Application Research of Measuring the Length of Gauge Blocks by Laser Interferometry

Huang Lei*, Ma Hong, and Xu Wenhui
Author Affiliations
  • [in Chinese]
  • show less

    The influences of gauge blocks gestures on calibration of coordinate measuring machine attach increasing attention. In particular, the effects of gauge blocks gestures on metrology coordinate measuring machine of which the measurement accuracy is hyper-0.6 μm+1000/L cannot be ingnored. Adopting laser interferometer as measuring datum, length deflection of gauge blocks are measured in different gestures by developing an adjustable platform. An advance multi-point bracing means is proposed to minish length variation of gauge blocks. The results of experiments indicate that the maximum length deflection of gauge blocks is +0.21 μm by adopting multi-point bracing means. According to the formula of error uncertainty for coordinate measuring machine the standard uncertainty is recalculated to be 0.75 μm.

    Tools

    Get Citation

    Copy Citation Text

    Huang Lei, Ma Hong, Xu Wenhui. Application Research of Measuring the Length of Gauge Blocks by Laser Interferometry[J]. Acta Optica Sinica, 2011, 31(s1): 100501

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: OPTOELECTRONICS

    Received: Mar. 5, 2011

    Accepted: --

    Published Online: Jun. 23, 2011

    The Author Email: Lei Huang (huanglei_mail@yahoo.com.cn)

    DOI:10.3788/aos201131.s100501