Spectroscopy and Spectral Analysis, Volume. 31, Issue 7, 2007(2011)

A Dual-Wavelength Spectroscopic Method for the Low Chemical Oxygen Demand Determination

JIANG Ran1、*, CHAI Xin-sheng2, ZHANG Cui2, and TANG Hong-liang1,3
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less

    The present paper reports a dual wavelength spectroscopic method for the determination of chemical oxygen demand(COD) of low content samples, based on the absorption measurement at 440 and 560 nm. Since the absorptivities of Cr3+ at these two wavelengths are the same, a net absorption from dichromate ions can be determined by subtracting the absorption at 560 nm(only contributed by Cr3+) from the absorption at 440 nm. The results showed that the detection limit of quantification is 8.6 mg·L-1 and the relative standard deviations are in the 2%~15% range for the low COD standard samples (≤100 mg·L-1) in this proposed method. Further more, the method does not require calibration based on the standard samples. The present method is simple, reliable, and accurate; it is suitable for the application in a mass testing for low COD samples.

    Tools

    Get Citation

    Copy Citation Text

    JIANG Ran, CHAI Xin-sheng, ZHANG Cui, TANG Hong-liang. A Dual-Wavelength Spectroscopic Method for the Low Chemical Oxygen Demand Determination[J]. Spectroscopy and Spectral Analysis, 2011, 31(7): 2007

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Sep. 18, 2010

    Accepted: --

    Published Online: Aug. 29, 2011

    The Author Email: Ran JIANG (xschai@gmail.com)

    DOI:10.3964/j.issn.1000-0593(2011)07-2007-04