Spectroscopy and Spectral Analysis, Volume. 30, Issue 12, 3268(2010)
Detection of Dissolved Organic Matter Based on Three-Dimensional First-Order Derivative Fluorescence Spectrometry
Based on three-dimensional first-order derivative fluorescence spectrometry, an analysis method for detecting dissolved organ matter in water is proposed in the present paper. By using simplified least squares differentiation methods presented by Savitzky and Goly, the first-order partial derivatives for emission wavelength and excitation wavelength were calculated. As the fitting polynomial has the smoothing function in the calculation of derivative spectra, a separate smoothing method is not required to remove spectrometry noise. The regression model was calculated by partial least square for 4-dimension fluorescence data including emission wavelength, excitation wavelength and their first-order derivatives. The Experimental results for detecting total organic carbon (TOC) in water show that the proposed method has obvious advantage over the conventional fluorescence spectrometry analysis methods in the aspect of the root mean square error of prediction and correlation coefficient.
Get Citation
Copy Citation Text
DU Shu-xin, DU Yang-feng, WU Xiao-li. Detection of Dissolved Organic Matter Based on Three-Dimensional First-Order Derivative Fluorescence Spectrometry[J]. Spectroscopy and Spectral Analysis, 2010, 30(12): 3268
Received: Jan. 26, 2010
Accepted: --
Published Online: Jan. 26, 2011
The Author Email: Shu-xin DU (shxdu@iipc.zju.edu.cn)
CSTR:32186.14.