Acta Optica Sinica, Volume. 42, Issue 4, 0404001(2022)
Responsivity Calibration Technology for Pyroelectric Terahertz Detector
In order to ensure the accuracy and reliability of terahertz radiation intensity measurement, the responsivity calibration of pyroelectric terahertz detectors is carried out. First, a terahertz detector responsivity calibration device is constructed based on the substitution method, and the combined standard uncertainty of the device is 2.4%. Then, the 12D-3S-VP type terahertz detector is calibrated by the calibration device at the frequency point of 1.63 THz, and the responsivity calibration result is 197.6 mV/W, which is consistent with the value given by the manufacturer. The self-developed pyroelectric terahertz detector is calibrated, and the responsivity and the combined standard measurement uncertainty are 362.2 mV/W and 2.7% respectively. Finally, the standard deviation of the calibration result is compared with the combined standard uncertainty, and it is found that the standard deviation is in a reasonable range, which further proves the rationality of the calibration method, calibration results and uncertainty analysis.