    Large-area plastic scintillators are widely used in radiation monitoring. They are typically employed as gamma radiation counters through signal over-threshold detection. The value of the threshold voltage affects the detector efficiency and minimum detectable activity. When rays are incident at different positions of the large-area plastic, the photon collection efficiencies generated through collecting end pair differ, leading to differences in the detector efficiency.

    Purpose

    This study aims to explore appropriate threshold voltage to reduce the differences and maintain a low minimum detectable activity (MDA), hence achieving high detection efficiency for large-area plastic scintillator.

    Methods

    Firstly, an energy spectra acquisition system was designed using STM32F429 produced by STM semiconductor as the main control chip and a counter for detector efficiency testing of large-area plastic scintillator in size of 400 mm×300 mm×50 mm. Then, the reasonable threshold voltage was determined according to the background energy spectra of the large-area plastic scintillator, the energy spectrum of the 60Co and 137Cs sources at different positions of the large-area plastic scintillator with acquisition time of 3 min, and signal-amplification relationship.

    Results

    The determined threshold voltage using above mentioned method for 60Co source is 93.7 mV. To achieve high detector efficiency and low MDA for the large-area plastic scintillator, the best determined threshold voltage for 60Co source is 95 mV. At this voltage, the detector efficiency for 60Co source is 22%, the MDA is 78 Bq.

    Conclusions

    The proposed method has reference value for large-area plastic scintillation designing and manufacturing detectors.

    Tools

    Get Citation

    Copy Citation Text

    Jiutao CHEN, Yushou SONG, Rui XU, Runzhou SONG, Shengqiang CHEN. Appropriate threshold voltage toward high detection efficiency of large-area plastic scintillator[J]. NUCLEAR TECHNIQUES, 2023, 46(12): 120403

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: May. 4, 2023

    Accepted: --

    Published Online: Mar. 7, 2024

    The Author Email: SONG Yushou (宋玉收)

    DOI:10.11889/j.0253-3219.2023.hjs.46.120403

    Topics