Acta Optica Sinica, Volume. 32, Issue 5, 528002(2012)

Study on Spectral Calibration of a Novel Interferometric Spectrometer with Supper-Resolution Capability

Shi Hailiang*, Fang Yonghua, Wu Jun, and Xiong Wei
Author Affiliations
  • [in Chinese]
  • show less

    In connection with the new spatial heterodyne spectroscopy (SHS), integrated grating diffraction and space interference, a study on the spectral calibration technology is done. From interference mechanism of SHS, the spectral calibration difference between SHS and traditional dispersive spectrometer is analyzed, study on ILS, spectral resolution, bandwidth calibration principle and method of SHS is done, and spectral calibration method of tunable laser with integrating sphere is designed and calibration data processing algorithm according to hyper-spectral data features is explored. Finally, using the above method CO2 spatial heterodyne spectrometer is calibrated experimentally, and calibration accuracy is validated using magnesium lamp. The results show that spectral calibration method can meet the requirements of SHS spectral calibration, and are consistent with theoretical values of prototype.

    Tools

    Get Citation

    Copy Citation Text

    Shi Hailiang, Fang Yonghua, Wu Jun, Xiong Wei. Study on Spectral Calibration of a Novel Interferometric Spectrometer with Supper-Resolution Capability[J]. Acta Optica Sinica, 2012, 32(5): 528002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Remote Sensing and Sensors

    Received: Oct. 13, 2011

    Accepted: --

    Published Online: Apr. 16, 2012

    The Author Email: Hailiang Shi (hlshi@aiofm.ac.cn)

    DOI:10.3788/aos201232.0528002