Acta Optica Sinica, Volume. 30, Issue 5, 1337(2010)
Advanced Calibration Method for Sun Radiometers
A CIMEL CE318 is calibrated by using a new tunable laser-based facility for spectral radiance responsivity with combined standard uncertainties lower than 1%. The same CE318 is also calibrated by using a NASA lamp-illuminated integrating sphere source. Relative deviations between NASA and our calibration coefficients are within ±1.4%;which indicates that the proposed method is accurate.
Get Citation
Copy Citation Text
Xu Qiuyun, Zheng Xiaobing, Zhang Wei, Wang Xianhua. Advanced Calibration Method for Sun Radiometers[J]. Acta Optica Sinica, 2010, 30(5): 1337
Category: Instrumentation, Measurement and Metrology
Received: May. 22, 2009
Accepted: --
Published Online: May. 11, 2010
The Author Email: Qiuyun Xu (qyxu@aiofm.ac.cn)