Journal of Atmospheric and Environmental Optics, Volume. 6, Issue 2, 122(2011)
Technology of Precise Measurement of Slit Aperture Area
In order to achieve the absolute uncertainty which is less than 0.05% and the relative stability which is 0.03% per year in 0.4~1.0 μm of solar absolute spectral irradiance observation, the slit aperture area must be measured with high-precision in the development of solar absolute irradiance spectra-radiometer. The flux comparison method is developed by using laser scanning for precise measurement of the slit aperture area. The laser-illuminated integrating sphere system is used as a light source, and the high-precision trap detector is used to receive signals. Small size apertures can be measured without a reference aperture. In the measurement of the entrance area of solar absolute irradiance spectra-radiometer of which the nominal area is 7 mm×0.1 mm, the total relative uncertainty is about 4.5×10-4.
Get Citation
Copy Citation Text
HE Bo, ZHENG Xiao-bing, LI Xin, LI Kai, XU Wen-bin. Technology of Precise Measurement of Slit Aperture Area[J]. Journal of Atmospheric and Environmental Optics, 2011, 6(2): 122
Category:
Received: Jun. 21, 2010
Accepted: --
Published Online: Mar. 28, 2011
The Author Email: Bo HE (herbird@sina.com)
CSTR:32186.14.