Optics and Precision Engineering, Volume. 25, Issue 11, 2845(2017)

Absolute measurement of photon flux for synchrotron radiation hard X-rays

LI Hua-peng1...2,*, ZHENG Wei-ning3, ZHAO Yi-dong1, WANG Pei-wei4 and LI Fan1 |Show fewer author(s)
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • 4[in Chinese]
  • show less

    The photon flux of the synchrotron radiation hard X-rays was measured in Beijing Synchrotron Radiation Facility (BSRF) using a free-air ionization chamber. Absolute air-kerma can be obtained with the free-air ionization chamber. The photon flux was calculated based on the relation between air-kerma and flux in the fundamental quantities and units for ionizing radiation. Because the photon flux changes over time during the measurement, an online monitoring system was applied, and the uncertainty of which is 0.03%. The air attenuation correction factor and ion recombination correction factor were measured at 15 keV synchrotron radiation X-rays. The measurement results show that the absolute air-kerma rate at 15 keV synchrotron radiation X-rays is 0.239 Gy/s, accordingly, the photon flux of the source is 5.28×109 photons/s, the relative uncertainty is 0.68%.

    Tools

    Get Citation

    Copy Citation Text

    LI Hua-peng, ZHENG Wei-ning, ZHAO Yi-dong, WANG Pei-wei, LI Fan. Absolute measurement of photon flux for synchrotron radiation hard X-rays[J]. Optics and Precision Engineering, 2017, 25(11): 2845

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Mar. 20, 2017

    Accepted: --

    Published Online: Jan. 17, 2018

    The Author Email: Hua-peng LI (hplee@ihep.ac.cn)

    DOI:10.3788/ope.20172511.2845

    Topics