Acta Optica Sinica, Volume. 33, Issue 12, 1229001(2013)

Optimal Fitting Cumulants Method for Dynamic Light Scattering

Liu Wei*, Wang Yajing, and Shen Jin
Author Affiliations
  • [in Chinese]
  • show less

    Dynamic light scattering is a powerful tool for measuring the size of nanoscale particles. However, the average particle size inverted by the traditional linear cumulants method depends closely on the length of correlation data. Linear and non-linear fitting algorithms are analyzed in order to overcome this disadvantage. An optimized fitting algorithm for the cumulants method is proposed based on the advantages of both fitting algorithms. The particle diameter is obtained from a first-order curve fit, and the particle distribution′s polydispersity from a second-order polynomial fit over the optimal range of the intensity correlation function. Theoretical analysis and experimental data show that the relative error and repeatability of the inverted diameter are less than 2%, and the relative error of polydispersity index is less than 6%. In conclusion, the optimal fitting algorithm for the cumulants method can be used to measure a stable and reliable particle size and its polydispersity index.

    Tools

    Get Citation

    Copy Citation Text

    Liu Wei, Wang Yajing, Shen Jin. Optimal Fitting Cumulants Method for Dynamic Light Scattering[J]. Acta Optica Sinica, 2013, 33(12): 1229001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Scattering

    Received: Jun. 25, 2013

    Accepted: --

    Published Online: Dec. 5, 2013

    The Author Email: Wei Liu (weikey@sdut.edu.cn)

    DOI:10.3788/aos201333.1229001