Spectroscopy and Spectral Analysis, Volume. 32, Issue 5, 1397(2012)

Preconcentration of Ultra-Trace Germanium in Water Samples with Nano-Sized TiO2 Colloid and Determination by HGAFS with Colloid Sampling

QIAN Sha-hua*, LU Min, and ZHANG Xu
Author Affiliations
  • [in Chinese]
  • show less

    A novel method for preconcentration of ultra-trace germanium and determination by hydride generation atomic fluorescence spectrometry (HGAFS) was presented in this work. When the pH values of solution were 6.0~8.0, the adsorption efficiency of nano-sized TiO2 colloid for germanium was 97.0%~99.0% in a short time. Nano-sized TiO2 colloid was used to concentrate the ultra-trace germanium in water sample due to its high absorption capacity. After centrifugation, the supernatant fluid was removed. The sediment which contained the concentrated Ge(Ⅳ) was inverted to colloid by adding HCl of certain concentration. Ge(Ⅳ) in colloid was directly determined by HGAFS. The proposed method possesses low detection limit (3σ) (0.060 μg·L-1) and good precision (the relative standard deviation (RSD) is 2.0%, n=6). The method is was also quite simple and time saving (without any filtration and desorption process). Satisfactory results were obtained when applying this method to the determination of ultra-trace germanium in water samples.

    Tools

    Get Citation

    Copy Citation Text

    QIAN Sha-hua, LU Min, ZHANG Xu. Preconcentration of Ultra-Trace Germanium in Water Samples with Nano-Sized TiO2 Colloid and Determination by HGAFS with Colloid Sampling[J]. Spectroscopy and Spectral Analysis, 2012, 32(5): 1397

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Nov. 24, 2011

    Accepted: --

    Published Online: Jun. 14, 2012

    The Author Email: Sha-hua QIAN (qiansh@whu.edu.cn)

    DOI:10.3964/j.issn.1000-0593(2012)05-1397-04