Spectroscopy and Spectral Analysis, Volume. 32, Issue 4, 1137(2012)

Modeling and Correction of the Effect of Detector Lateral Fringe Error in the Large Aperture Static Imaging Spectrometer

XIANGLI Bin1、*, L Qun-bo1, YUAN Yan2, HUANG Min1, ZHOU Jin-song1, and ZHANG Gui-feng1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    When the large aperture static imaging spectrometry (LASIS) was used in spaceborne remote sensing, the spectrum could not be well recovered from the interferogram with detector lateral fringe error. In order to obtain the more accurate spectrum, the error must be corrected. According to the imaging principle analysis of LASIS, the imaging model with lateral fringe error was given, and the error correction method was presented. Finally, the error correction method was validated with the data acquired by LASIS system. The results indicated that the error was commendably corrected by the method mentioned above, and the accuracy of recovered spectrum was evidently improved.

    Tools

    Get Citation

    Copy Citation Text

    XIANGLI Bin, L Qun-bo, YUAN Yan, HUANG Min, ZHOU Jin-song, ZHANG Gui-feng. Modeling and Correction of the Effect of Detector Lateral Fringe Error in the Large Aperture Static Imaging Spectrometer[J]. Spectroscopy and Spectral Analysis, 2012, 32(4): 1137

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Jul. 27, 2011

    Accepted: --

    Published Online: Apr. 16, 2012

    The Author Email: Bin XIANGLI (xiangli@aoe.ac.cn)