Spectroscopy and Spectral Analysis, Volume. 43, Issue 11, 3436(2023)

Determination of Mn, Co, Ni in Ternary Cathode Materials With Homologous Correction EDXRF Analysis

LIN Hong-jian1、*, ZHAI Juan1, LAI Wan-chang1, ZENG Chen-hao1,2, ZHAO Zi-qi1, SHI Jie1, and ZHOU Jin-ge1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Lithium-ion batteries play an important role in developing and applying new energy. The composition ratio of ternary cathode materials has a major impact on the performance and quality of lithium-ion battery products, and production control requires timely and accurate control of composition changes in the mix. Energy dispersive X-ray fluorescence (EDXRF) technology has a good prospect for rapid analysis in this area, but the analytical accuracy of commercial instruments cannot currently meet production requirements. To solve the technical problem of high-precision EDXRF analysis of ternary cathode material composition Mn, Co, Ni, a self-calibrating formal EDXRF analysis technology based on homologous excitation is proposed in this paper, which uses a tungsten target X-ray tube (25 kV/400 μA) and two electrically cooled SDD detectors with the energy resolution of 135 eV (@5.9 keV) to form a two-channel synchronous X-ray fluorescence excitation and detection device. After splitting the primary X-rays emitted by the X-ray tube by the dual channel collimator, the calibration sample and the sample to be measured are excited. The two detectors simultaneously measure the fluorescence counts of the two samples and use the energy spectrum data of the standard sample to perform “normalization” processing to achieve synchronous correction of the energy spectrum data of the sample to be tested, thus reducing the influence of X-ray tube instability in the analytical instrument. The stability instrument was analyzed regarding the count decay rate, count variation, and overall effect through 140 repeatability tests in 8 hours and compared with that of the single optical path. The relative standard deviation and maximum relative deviation were used as evaluation indices to assess the stability instrument. The counting attenuation rate decreased from -0.049 3%·h-1 of the single optical path to 0.001 0%. For 11 data points with large fluctuations, the relative standard deviation decreased from 0.151 4% to 0.032 6% of the single optical path, indicating that the self-calibrating formal EDXRF analysis technology of homologous excitation can effectively reduce the effects of counting attenuation and primary X-ray energy spectrum fluctuations. From the perspective of comprehensive effect, the relative standard deviation and maximum relative deviation of Mn, Co, and Ni are 0.076% and 0.170%, respectively, after the correction of synchronous data, which is twice as stable as that of the single optical path. The paper establishes a mathematical model for quantitative analysis based on dual optical path EDXRF analysis. Through experimental verification, the absolute errors of Mn (17.361%~20.016%), Co (12.991%~14.965%), Ni (29.653%~33.065%) in powder compacted samples are not more than -0.072%, -0.061%, 0.098%, respectively, and the analysis time of single sample is 200 s, indicating that the self-calibration formal EDXRF analysis technology with homologous excitation can effectively improve the instrument analysis accuracy, and achieve fast accurate testing requirements.

    Tools

    Get Citation

    Copy Citation Text

    LIN Hong-jian, ZHAI Juan, LAI Wan-chang, ZENG Chen-hao, ZHAO Zi-qi, SHI Jie, ZHOU Jin-ge. Determination of Mn, Co, Ni in Ternary Cathode Materials With Homologous Correction EDXRF Analysis[J]. Spectroscopy and Spectral Analysis, 2023, 43(11): 3436

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Mar. 10, 2023

    Accepted: --

    Published Online: Nov. 26, 2023

    The Author Email: Hong-jian LIN (489469519@qq.com)

    DOI:10.3964/j.issn.1000-0593(2023)11-3436-09

    Topics