Chinese Optics Letters, Volume. 11, Issue s2, S21701(2013)

Investigation of glucose concentration measurement based on tunable pulsed laser induced photoacoustic technique

Zhong Ren, Guodong Liu, and Zhen Huang

A tunable pulsed laser induced photoacoustic (PA) measurement set-up is established in the forward mode to monitor in vitro glucose concentration. A series of experiments are investigated to verify the feasibility of this set-up and scheme. Peak-to-peak values (PPVs) of several glucose aqueous solutions are recorded and averaged 512 times at each wavelength. Experimental results demonstrate that the time-resolved PA profile of glucose solutions has a good agreement with the PA theories. The characteristic wavelengths of glucose solution are determined via differential method. The root-mean-square error (RMSE) of predicted concentration reaches 3.15 mg/dl at the optimum wavelength of 1 510 nm via least square (LS) fitting algorithm.

Tools

Get Citation

Copy Citation Text

Zhong Ren, Guodong Liu, Zhen Huang. Investigation of glucose concentration measurement based on tunable pulsed laser induced photoacoustic technique[J]. Chinese Optics Letters, 2013, 11(s2): S21701

Download Citation

EndNote(RIS)BibTexPlain Text
Save article for my favorites
Paper Information

Category: Medical optics and biotechnology

Received: Jan. 30, 2013

Accepted: Mar. 8, 2013

Published Online: Jul. 22, 2013

The Author Email:

DOI:10.3788/col201311.s21701

Topics