Spectroscopy and Spectral Analysis, Volume. 31, Issue 5, 1375(2011)

Design of Full-Polarized and Multi-Spectral Imaging System Based on LCVR

ZHANG Ying*, ZHAO Hui-jie, CHENG Xuan, and XIONG Sheng-jun
Author Affiliations
  • [in Chinese]
  • show less

    A new full-polarized multi-spectral imaging system is described, which uses electronically controlled LCVR (liquid crystal variable retarder) to modulate the full-polarized state of light in the visible to IR range. The system consisted of optical lenses, LCVRs, filters and CCD. Firstly, the system structure, working theory and optical design are introduced in the present paper. A polarization calibration method is provided and the calibration system was set up, which realized high-precision polarization calibration using a small polarized source. Then, a field experiment with the imaging system was carried out. Polarized spectral images with higher spectral and spatial resolution were collected. Finally, the data acquired were rough processed to get polarization degree image of the targets. It is concluded that the experiment has proved that the imaging system is effective in obtaining full-polarized and multi-spectral data. The image captured by the system can be applied to object identification and object classification.

    Tools

    Get Citation

    Copy Citation Text

    ZHANG Ying, ZHAO Hui-jie, CHENG Xuan, XIONG Sheng-jun. Design of Full-Polarized and Multi-Spectral Imaging System Based on LCVR[J]. Spectroscopy and Spectral Analysis, 2011, 31(5): 1375

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Jul. 27, 2010

    Accepted: --

    Published Online: May. 30, 2011

    The Author Email: Ying ZHANG (yingzhang@buaa.edu.cn)

    DOI:

    Topics