Laser & Optoelectronics Progress, Volume. 53, Issue 7, 70401(2016)

Study on Thermoelectrically Cooled CCD in Raman Spectrum Detection System for Gas Analysis

Zhao Jun*, Zuo Duluo, and Wang Xinbing
Author Affiliations
  • [in Chinese]
  • show less

    The commonly used CCD in Raman spectroscopy is sensitive to temperature, and cooling can effectively reduce its dark current noise. A cooling system is designed, and the finite element analysis software ANSYS is used to simulate the heat conduction of the heat sink and the heat exchange of gas flow in the vacuum chamber. The temperature change with the cooling time is measured at different vacuum degrees and with different buffer gas including Ar, Ne, N2 and air. The experimental results show that better cooling effect can be realized when N2 is used as buffer gas, and higher vacuum degree does not guarantee better cooling effect, which is consistent with the thermal simulation result. The measured dark current noise of CCD at different temperatures shows that CCD temperature has strong impact on its dark current noise, cooling can reduce the noise effectively, and the noise varies with temperature exponentially.

    Tools

    Get Citation

    Copy Citation Text

    Zhao Jun, Zuo Duluo, Wang Xinbing. Study on Thermoelectrically Cooled CCD in Raman Spectrum Detection System for Gas Analysis[J]. Laser & Optoelectronics Progress, 2016, 53(7): 70401

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Detectors

    Received: Jan. 25, 2016

    Accepted: --

    Published Online: Jul. 8, 2016

    The Author Email: Jun Zhao (525418495@qq.com)

    DOI:10.3788/lop53.070401