Journal of Innovative Optical Health Sciences, Volume. 8, Issue 4, 1550022(2015)

Constrained regularization for noninvasive glucose sensing using Raman spectroscopy

Wei-Chuan Shih*
Author Affiliations
  • Department of Electrical and Computer Engineering Department of Biomedical Engineering, University of Houston 4800 Calhoun Rd., Houston, TX 77204, USA
  • show less

    Multivariate calibration is an important tool for spectroscopic measurement of analyte concentrations. We present a detailed study of a hybrid multivariate calibration technique, constrained regularization (CR), and demonstrate its utility in noninvasive glucose sensing using Raman spectroscopy. Similar to partial least squares (PLS) and principal component regression (PCR), CR builds an implicit model and requires knowledge only of the concentrations of the analyte of interest. Calibration is treated as an inverse problem in which an optimal balance between model complexity and noise rejection is achieved. Prior information is included in the form of a spectroscopic constraint that can be obtained conveniently. When used with an appropriate constraint, CR provides a better calibration model compared to PLS in both numerical and experimental studies.

    Tools

    Get Citation

    Copy Citation Text

    Wei-Chuan Shih. Constrained regularization for noninvasive glucose sensing using Raman spectroscopy[J]. Journal of Innovative Optical Health Sciences, 2015, 8(4): 1550022

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Sep. 13, 2014

    Accepted: Nov. 24, 2014

    Published Online: Jan. 10, 2019

    The Author Email: Shih Wei-Chuan (wshih@uh.edu)

    DOI:10.1142/s1793545815500224

    Topics