Acta Photonica Sinica, Volume. 35, Issue 6, 881(2006)
Imaging Research with Non-periodic Multilayers
A grazing KB microscope was designed for hard X-ray (8 keV;Cu Ka radiation) imaging in ICF diagnostic experiments. Ray tracing software (ZEMAX) is used to simulate optical system performance. The best theoretical resolution of KB microscope is about 2 micron and better than 10 micron in 200 micron field of view. W/B4C supermirror was deposited on the Si wafer substrate and the reflectivity was measured by X-ray diffraction instrument (XRD). The reflectivity of supermirror is about 20% in 0.3°bandwidth. 8keV Cu target X-ray tube source was used in x-ray imaging experiments and the magnification of 1× and 2× X-ray images are obtained.
Get Citation
Copy Citation Text
Gu Chunshi, Wang Zhanshan, Mu Baozhong, Wang Fengli, Zhang Zhong, Qin Shuji, Chen Lingyan. Imaging Research with Non-periodic Multilayers[J]. Acta Photonica Sinica, 2006, 35(6): 881
Category:
Received: Feb. 20, 2006
Accepted: --
Published Online: Jun. 3, 2010
The Author Email: Zhanshan Wang (wangzs@mail.tongji.edu.cn)