Acta Optica Sinica, Volume. 31, Issue 12, 1212002(2011)
A New Background Radiation-Balanced Model for the Size-of-Source Effect Measurement
The background radiation effect of the measurement for the size-of-source effect (SSE) of radiation thermometers must be considered at low and median temperature. The calculation model includes items of radiation thermometer outputs to measure an infinite-size radiation source. After analyzing the influence of the detector temperature on the models of temperature and SSE measurements, a novel calculation model, utilizing a suppositional detector temperature balanced with the background radiation for the SSE measurement is presented. This method prevents the need for an infinite-size radiation source and unavoidable approximation. The relations of the SSE and the increase of temperature reading are given. The measurement conditions of the source temperature, reading resolution and thermometer emissivity setting are analyzed to optimize the result of the SSE measurement.
Get Citation
Copy Citation Text
Yuan Zundong. A New Background Radiation-Balanced Model for the Size-of-Source Effect Measurement[J]. Acta Optica Sinica, 2011, 31(12): 1212002
Category: Instrumentation, Measurement and Metrology
Received: May. 17, 2011
Accepted: --
Published Online: Oct. 31, 2011
The Author Email: Zundong Yuan (yuanzd@nim.ac.cn)