Journal of Terahertz Science and Electronic Information Technology , Volume. 22, Issue 5, 523(2024)

Test and analysis of electromagnetic compatibility of heart sound electrocardiograph detector and research on rectification method

WANG Shuotong1... SHI Pengcheng1, ZHANG Guojun1,*, CUI Jiangong1, LI Yirui1, WANG Bo1 and WANG Tao2 |Show fewer author(s)
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Modern medical devices are more inclined to high precision, high sensitivity, and miniaturization, resulting in more and more complex electromagnetic environments. External electromagnetic interference will also affect medical devices, and even affect doctors' diagnosis in severe cases. In this paper, combined with the heart sound electrocardiograph detector, the conduction emission test, radiation emission test and electrostatic discharge test are studied based on the medical device test standard. Some rectification methods are proposed, such as rectifying the shell, adding the filter circuit, adding the magnetic ring and adding the charge release circuit. It can be used as a reference for the analysis and rectification of other types of related medical devices.

    Tools

    Get Citation

    Copy Citation Text

    WANG Shuotong, SHI Pengcheng, ZHANG Guojun, CUI Jiangong, LI Yirui, WANG Bo, WANG Tao. Test and analysis of electromagnetic compatibility of heart sound electrocardiograph detector and research on rectification method[J]. Journal of Terahertz Science and Electronic Information Technology , 2024, 22(5): 523

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Sep. 2, 2022

    Accepted: --

    Published Online: Aug. 22, 2024

    The Author Email: Guojun ZHANG (zhangguojun1977@nuc.edu.com)

    DOI:10.11805/tkyda2022162

    Topics