Acta Optica Sinica, Volume. 35, Issue 8, 812001(2015)

Optical Design and Stray Light Suppression of A New Portable Echelle Spectrometer

Yang Jin1,2、*, Yin Lu1,2, Yao Xuefeng1, Bayanheshig1, Tang Yuguo1, and Zhu Jiwei1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Echelle spectrometer is becoming a hot research and focus of modern spectrograph with the outstanding features of high resolution, small size, full spectrum transient direct- reading. In order to further refine its volume, improve the signal to noise ratio and the detection limit, a new structure of echelle spectrometer is designed. By folding main optical path, the new design has smaller size and lower stray light without reducing spectral resolution and image quality. It′ s optical size is less than 165 mm × 70 mm × 65 mm, and spectral resolution is 0.06 nm@200 nm. The baffle and diaphragm is designed to further reduce stray light. Through ray tracing simulation, the stray light of the new echelle spectrometer is less than 2×10-5 , the signal to noise ratio is improved significantly.

    Tools

    Get Citation

    Copy Citation Text

    Yang Jin, Yin Lu, Yao Xuefeng, Bayanheshig, Tang Yuguo, Zhu Jiwei. Optical Design and Stray Light Suppression of A New Portable Echelle Spectrometer[J]. Acta Optica Sinica, 2015, 35(8): 812001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Nov. 24, 2014

    Accepted: --

    Published Online: Jul. 23, 2015

    The Author Email: Jin Yang (yangjinl@mail.ustc.edu.cn)

    DOI:10.3788/aos201535.0812001

    Topics