Acta Optica Sinica, Volume. 39, Issue 7, 0714002(2019)
Absolute Radiance Responsivity Calibration Based on Supercontinuum Laser and Monochromator
A spectrally resolved scanning calibration facility based on a supercontinuum laser and monochromator (SCM) was proposed for the calibration of the absolute spectral radiance responsivity of the sensors. Two radiance meters, labeled Traps-A and Traps-B, were used herein. The absolute spectral power responsivities of the meters were traced to a cryogenic absolute radiometer. The absolute spectral radiance responsivity of Trap-B was obtained by a general unit-level calibration method, and a system-level calibration method based on the SCM with Trap-A taken as the reference. The calibration uncertainties of Traps-A and Traps-B were lower than 0.46% and 1.8%, respectively. The calibration results by the two methods were in good agreement, with a relative difference of less than 0.9% in the 450-900 nm. The results imply that a spectrally-resolved scanning calibration facility based on SCM is suitable for the absolute spectral radiometric calibration of sensors and highly applicable to the absolute spectral radiometric calibration of remote sensors.
Get Citation
Copy Citation Text
Chenyu Xie, Wenchao Zhai, Jianjun Li, Fang Gao, Yue Li, Haoyu Wu, Xiaobing Zheng. Absolute Radiance Responsivity Calibration Based on Supercontinuum Laser and Monochromator[J]. Acta Optica Sinica, 2019, 39(7): 0714002
Category: Lasers and Laser Optics
Received: Dec. 20, 2018
Accepted: Mar. 25, 2019
Published Online: Jul. 16, 2019