Acta Optica Sinica, Volume. 38, Issue 4, 0429001(2018)

Measurement Upper Limit of Laser Particle Size Analyzer

Hua Hu1,2, Fugen Zhang3、*, Qieni Lü1,2, and Linchao Pan1,2
Author Affiliations
  • 1 School of Precision Instrument and Opto-Electronics Engineering, Tianjin University, Tianjin 300072, China
  • 2 Key Laboratory of Opto-Electronics Information Technology of Ministry of Education, Tianjin 300072, China
  • 3 Zhuhai Linkoptik Instruments Co., Ltd., Zhuhai, Guangdong 519085, China
  • show less

    The singular value decomposition method is used to study the characteristics of light energy coefficient matrix of a laser particle size analyzer, and the sensitivity parameters are defined to express the variance of particle size distribution with the light energy distribution. The measurement upper limit under certain parameters is obtained. An analytic formula which discloses the relationship between the measurement upper limit and the physical parameters is derived, which shows that the measurement upper limit of the instrument is determined by the minimum acceptance angle of this instrument for the scattered light under the condition of constant incident wavelength. The experimental results verify the correctness of this formula.

    Tools

    Get Citation

    Copy Citation Text

    Hua Hu, Fugen Zhang, Qieni Lü, Linchao Pan. Measurement Upper Limit of Laser Particle Size Analyzer[J]. Acta Optica Sinica, 2018, 38(4): 0429001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Scattering

    Received: Sep. 28, 2017

    Accepted: --

    Published Online: Jul. 10, 2018

    The Author Email: Zhang Fugen (zhang@linkoptik.com)

    DOI:10.3788/AOS201838.0429001