  • show less

    Reference sample was prepared by masterbatch method, containing heavy metals with known concentration of electronic information products (plastic), the repeatability and precision were determined, and reference sample preparation procedures were established. X-Ray fluorescence spectroscopy (XRF) analysis method was used to determine the repeatability and uncertainty in the analysis of the sample of heavy metals and bromine element. The working curve and the metrical methods for the reference sample were carried out. The results showed that the use of the method in the 200~2 000 mg·kg-1 concentration range for Hg, Pb, Cr and Br elements, and in the 20~200 mg·kg-1 range for Cd elements, exhibited a very good linear relationship, and the repeatability of analysis methods for six times is good. In testing the circuit board ICB288G and ICB288 from the Mitsubishi Heavy Industry Company, results agreed with the recommended values.

    Tools

    Get Citation

    Copy Citation Text

    YAO Wen-qing, ZONG Rui-long, ZHU Yong-fa. Standard Sample Preparation Method for Quick Determination of Trace Elements in Plastic[J]. Spectroscopy and Spectral Analysis, 2011, 31(8): 2274

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Sep. 15, 2010

    Accepted: --

    Published Online: Aug. 29, 2011

    The Author Email: Wen-qing YAO (yaowq@tsinghua.edu.cn)

    DOI:10.3964/j.issn.1000-0593(2011)08-2274-04

    Topics