Acta Optica Sinica, Volume. 21, Issue 2, 232(2001)

Precision Angle Sensor and Its Calibration Techniques

[in Chinese]1, [in Chinese]1, [in Chinese]1, [in Chinese]2, and [in Chinese]2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    A high-precision angle sensor and a new method for calibrating precision angle sensor absolutely without any accurate reference instruments are described. Two angle sensors with compensate devices which compensating thermal drift of the light source are developed and calibrated by themselves. The measurement ranges of two angle sensors are 180″, and the resolutions of two angle sensors are 0.003″ and 0.004″, respectively. The repeatability error of the evaluated mean sensitivity was improved to 0.015% by compensating device. The repeatability of the calibrated linearity error was reduced to 0.015″.

    Tools

    Get Citation

    Copy Citation Text

    [in Chinese], [in Chinese], [in Chinese], [in Chinese], [in Chinese]. Precision Angle Sensor and Its Calibration Techniques[J]. Acta Optica Sinica, 2001, 21(2): 232

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Devices

    Received: Sep. 30, 1999

    Accepted: --

    Published Online: Aug. 10, 2006

    The Author Email:

    DOI:

    Topics