INFRARED, Volume. 41, Issue 5, 30(2020)

Research on Measurement Method of Infrared Thermal Imager for Compensating High Temperature Effect

Du-min LEI*
Author Affiliations
  • [in Chinese]
  • show less

    Aiming at the problem of inaccurate temperature measurement of the common infrared thermal imager caused by the thermal radiation of the surrounding high temperature objects, according to the principle of the temperature measurement and the infrared thermal radiation theory, a high temperature compensation measurement method based on this type of infrared thermal imager is designed. Shielding materials with emissivity of 0.5 can fully compensate the effect of the high temperature radiation. When the emissivity is equal to 1, it is equivalent to ignoring the thermal radiation of the high temperature object, which does not play a compensating role. The higher the emissivity is above 0.5, the higher the measured temperature is than the real temperature; and the more 0.5 exceeds the emissivity, the higher the real temperature is than the measured temperature.

    Tools

    Get Citation

    Copy Citation Text

    LEI Du-min. Research on Measurement Method of Infrared Thermal Imager for Compensating High Temperature Effect[J]. INFRARED, 2020, 41(5): 30

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Apr. 28, 2020

    Accepted: --

    Published Online: Nov. 11, 2020

    The Author Email: Du-min LEI (2934493662@qq.com)

    DOI:10.3969/j.issn.1672-8785.2020.05.005

    Topics