Acta Photonica Sinica, Volume. 45, Issue 7, 70706001(2016)
Multi-probe Fiber Thermometer Based on Sapphire-ruby Fiber Sensing Head
To measure the temperature range from room temperature to 1 000℃, a sapphire-ruby fiber -based multi-probe thermometer was proposed. The thermometer system combines fluorescence lifetime with radiationand, which can measure the temperature of 8 probes at the same time. A digital iteration algorithm based on area balance was presented to measure the fluorescence lifetime. Compared with the Fast Fourier Transformation algorithm through simulation, the proposed algorithm features robust, high efficiency, strong anti-noise and immune to background level. The temperature test and stability test of the system were studied for a long time, the results show that the system has high stability and accuracy, the measured temperature range from room temperature to 1 000℃, the accuracy can be ±1℃.
Get Citation
Copy Citation Text
SU Jian-jia, JIANG Pei-pei, CUI Wei, WU Bo, SHEN Yong-hang. Multi-probe Fiber Thermometer Based on Sapphire-ruby Fiber Sensing Head[J]. Acta Photonica Sinica, 2016, 45(7): 70706001
Received: Dec. 16, 2015
Accepted: --
Published Online: Aug. 18, 2016
The Author Email: Jian-jia SU (sujianjia90@163.com)