Acta Optica Sinica, Volume. 31, Issue 8, 812007(2011)

Evaluation of Incident Light Intensity Based on Gregory Theory of Light Fluctuation for In-Line Particle Sizing Measurement

[in Chinese]*, [in Chinese], [in Chinese], and [in Chinese]
Author Affiliations
  • [in Chinese]
  • show less

    For in-line measurement of particle size and concentration based on the light fluctuation method, the intensity of incident light must be pre-detected. However, it is very difficult to measure the incident light intensity during the in-line measurement because the incident light intensity is easily affected by many factors. A novel method based on Gregory′s light fluctuation method is proposed. The relation among the intensity of incident light, the ratio of particle concentrations at different time, the standard deviation and the mean value of the intensity of transmitted light is established. With those measurable parameters the intensity of incident light may be obtained. Experimental results agree well with the pre-detected value of the intensity of incident light, and the maximum deviation is less than 1.1%.

    Tools

    Get Citation

    Copy Citation Text

    [in Chinese], [in Chinese], [in Chinese], [in Chinese]. Evaluation of Incident Light Intensity Based on Gregory Theory of Light Fluctuation for In-Line Particle Sizing Measurement[J]. Acta Optica Sinica, 2011, 31(8): 812007

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Jan. 25, 2011

    Accepted: --

    Published Online: Jul. 19, 2011

    The Author Email: (shaani@163.com)

    DOI:10.3788/aos201131.0812007

    Topics