Acta Photonica Sinica, Volume. 48, Issue 9, 912003(2019)

Field Calibration of Sunphotometer in Dunhuang Test Site

HUANG Dong1,2、*, LI Xin1, ZHANG Yanna1, ZHANG Quan1,2, ZHENG Xiaobing1, and SUN Ling3
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less

    In order to ensure the accuracy of aerosol optical characteristics observation in onorbit automatic calibration of satellites, field calibration techniques of sunphotometer in Dunhuang test site were studied, including Langley calibration based on relatively stable atmospheric conditions and cross calibration based on high precision reference instruments. The calibration results show that the maximum deviations of aerosol optical depth inversion from high precision sunphotometer compared with CE318 sunphotometer are within 0.015 and 0.01 respectively by Langley method and crosscalibration method. The effects of atmospheric conditions and instrument parameters on field calibration were calculated and analyzed. The calibrated high precision sunphotometer and CE318 sunphotometer were placed in Qinghai Lake for synchronous comparison observation. The maximum deviation of aerosol optical depth is within 0.01. The influence of field calibration error of sunphotometer in Dunhuang test site on onorbit calibration of satellite is less than 0.21%, which shows that the calibration results can meet the accuracy requirements of satellite onorbit automatic calibration.

    Tools

    Get Citation

    Copy Citation Text

    HUANG Dong, LI Xin, ZHANG Yanna, ZHANG Quan, ZHENG Xiaobing, SUN Ling. Field Calibration of Sunphotometer in Dunhuang Test Site[J]. Acta Photonica Sinica, 2019, 48(9): 912003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Jan. 22, 2019

    Accepted: --

    Published Online: Oct. 12, 2019

    The Author Email: Dong HUANG (hd125014@mail.ustc.edu.cn)

    DOI:10.3788/gzxb20194809.0912003

    Topics