Acta Photonica Sinica, Volume. 47, Issue 7, 718001(2018)
Experimental Study on Micro Quantitative Phase Imaging by Quadriwave Shearing Interferometry
A micro quantitative phase imaging method based on quadriwave shearing interferometry is proposed. A chessboard phase grating is used to get the interferogram of biological samples, and the phase information is obtained by the fast Fourier transform method. A pack of wavefront produced by the micro-electromechanical systems is measured, and the ZYGO interferometer is used for comparision. The error of phase measurement is not more than 3%, which verifies phase detection accuracy of the quadriwave shearing interferometer. A quantitative phase imaging system based on the quadriwave shearing interferometry is established, and a clear quantitative phase image is obtained from murine hepatoma cell. The experimental results show that the system based on the quadriwave shearing interferometry can achieve high-precision quantitative phase imaging, and is suitable for phase imaging of biological living cells.
Get Citation
Copy Citation Text
CUI Bo-chuan, WANG Jian-li, YAO Kai-nan, CHEN Tao. Experimental Study on Micro Quantitative Phase Imaging by Quadriwave Shearing Interferometry[J]. Acta Photonica Sinica, 2018, 47(7): 718001
Received: Jan. 29, 2018
Accepted: --
Published Online: Sep. 16, 2018