Optical Instruments, Volume. 37, Issue 2, 100(2015)

Linear test method of ultraviolet-visible spectrophotometer

LIU Minglan*... SUN Yuguo and ZHU Xuanxuan |Show fewer author(s)
Author Affiliations
  • School of Optical-Electrical and Computer Engineering,University of Shanghai for Science and Technology,Shanghai 200093,China
  • show less

    In order to qualitatively test linearity of ultraviolet-visible (UV-VIS) spectrophotometer, we test a food-dye absorbance at a certain wavelength using a UV-VIS spectrophotometer based on the water quality standard liquid calibration method, and then use a basic mathematical method to calculate volume concentration at each test point. We get a linear relationship of volume concentration and absorbance. Experiments show that this method is easy and convenient to measure the linearity of the instrument and the calculation method is accurate and easy to understand.

    Tools

    Get Citation

    Copy Citation Text

    LIU Minglan, SUN Yuguo, ZHU Xuanxuan. Linear test method of ultraviolet-visible spectrophotometer[J]. Optical Instruments, 2015, 37(2): 100

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Aug. 16, 2014

    Accepted: --

    Published Online: May. 20, 2015

    The Author Email: Minglan LIU (liuminglan665@126.com)

    DOI:10.3969/j.issn.1005-5630.2015.02.002