Spectroscopy and Spectral Analysis, Volume. 31, Issue 10, 2865(2011)

Analysis of Collimation Error in Fourier Transform Infrared Spectrometer Based on Step Mirrors

L Jin-guang1,2、*, LIANG Zhong-zhu1, FU Jian-guo1,2, and LIANG Jing-qiu1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    The collimation error in Fourier transform spectrometer has many effects on the interferogram and the spectrogram. By means of modulation, the authors found the reverse of contrast in the interferogram and the side slope noise in the spectrogram due to the collimation error. The authors analyzed the signal-to-noise ratio and resolution at different factor of beam divergence. When the factor of the beam divergence is 0.15°·mm-1, the signal-to-noise ratio can reduce to 6 dB and the resolution can go bad to 13.4 cm-1.

    Tools

    Get Citation

    Copy Citation Text

    L Jin-guang, LIANG Zhong-zhu, FU Jian-guo, LIANG Jing-qiu. Analysis of Collimation Error in Fourier Transform Infrared Spectrometer Based on Step Mirrors[J]. Spectroscopy and Spectral Analysis, 2011, 31(10): 2865

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Nov. 26, 2010

    Accepted: --

    Published Online: Nov. 9, 2011

    The Author Email: Jin-guang L (jinguanglv@163.com)