Spectroscopy and Spectral Analysis, Volume. 32, Issue 3, 817(2012)

FI-KR Non-Separated Method Coupled with FAAS for the Determination of Fe(Ⅱ) and Fe(Ⅲ) in Water

WANG Zhong-yuan1、*, ZHANG Hong-kang2, FANG Hong-da1, SU Yao-dong3, and Gauri S Mittal4
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • 4[in Chinese]
  • show less

    An FI-KR non-separated method coupled with FAAS for the determination of Fe(Ⅱ) and Fe(Ⅲ) was developed. With 60 s of sampling at a flow rate of 6.0 mL·min-1, EF of 41 for Fe(Ⅲ) and 9 for Fe(Ⅱ) were obtained. The precision(RSD, n=11) for Fe(Ⅲ) and Fe(Ⅱ) was 2.3% and 3.1% at the 0.04 mg·L-1 level respectively. When 0.1 ‰φ TEA was used as masking reagent, the recovery rate for Fe(Ⅲ) and Fe(Ⅱ) was from 97% to 101 % and from 96% to 100% respectively.

    Tools

    Get Citation

    Copy Citation Text

    WANG Zhong-yuan, ZHANG Hong-kang, FANG Hong-da, SU Yao-dong, Gauri S Mittal. FI-KR Non-Separated Method Coupled with FAAS for the Determination of Fe(Ⅱ) and Fe(Ⅲ) in Water[J]. Spectroscopy and Spectral Analysis, 2012, 32(3): 817

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Jun. 23, 2011

    Accepted: --

    Published Online: Apr. 16, 2012

    The Author Email: Zhong-yuan WANG (zhongyuan764@126.com)

    DOI:10.3964/j.issn.1000-0593(2012)03-0817-04