Spectroscopy and Spectral Analysis, Volume. 32, Issue 4, 1028(2012)

Research on the Method of Transient Spectrum Detection Based on Array CCD

ZHU Bing-li1,2、*, BAI Yong-lin3, WANG Bo3, LIU Bai-yu3, OUYANG Xian3, YANG Wen-zheng3, BAI Xiao-hong3, QIN Jun-jun3, ZHAO Jun-ping3, GOU Yong-sheng3, and LU Kai1,2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less

    Based on the characteristic of high speed line scanning for CCD in transient spectrum detection, a method of transient spectrum detection with array CCD is presented. The high speed line scanning with array CCD was realized by changing the mode of charge transfer. In order to explore the feasibility of this method, a fast detection system of single point based on linear CCD was designed and fabricated. Seven different LED pulses were measured when the system worked at fast detection mode of single point and normal mode respectively. The results demonstrate that the method of fast detection of single point based on linear CCD is feasible, and the rate of single point detection reaches up to 20 MHz. Thus, in theory, it was proved that transient spectrum detection with array CCD by changing the mode of charge transfer is also feasible.

    Tools

    Get Citation

    Copy Citation Text

    ZHU Bing-li, BAI Yong-lin, WANG Bo, LIU Bai-yu, OUYANG Xian, YANG Wen-zheng, BAI Xiao-hong, QIN Jun-jun, ZHAO Jun-ping, GOU Yong-sheng, LU Kai. Research on the Method of Transient Spectrum Detection Based on Array CCD[J]. Spectroscopy and Spectral Analysis, 2012, 32(4): 1028

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Aug. 23, 2011

    Accepted: --

    Published Online: Apr. 16, 2012

    The Author Email: Bing-li ZHU (zhubing_2006@163.com)

    DOI:10.3964/j.issn.1000-0593(2012)04-1028-04

    Topics