NUCLEAR TECHNIQUES, Volume. 46, Issue 10, 100402(2023)

Digital shaping methods for nuclear pulse signal based on cosine function

Yong WEI1,2, Huaiqiang ZHANG1,2、*, Yunchen QIAN1, and He CHEN1
Author Affiliations
  • 1Engineering Research Center of Nuclear Techniques Application, East China University of Technology, Ministry of Education, Nanchang 330013, China
  • 2School of Nuclear Science and Engineering, East China University of Technology, Nanchang 330013, China
  • show less

    Background

    Cosine function shaping (or cos shaping) is used to digitally shape nuclear pulse signals, as the shaping method is simple and has high operability and flexibility.

    Purpose

    This study aims to explore different cosine shaping methods of nuclear pulse signals, and evaluate their effect.

    Methods

    Firstly, based on the single exponential decay signal and cosine pulse signal, transfer functions and cascade formulas of three different cosine shaping methods in the Z-domain were derived. The influence of the parameter selection on the shaping effect in the cosine shaping algorithm was analyzed. Then, the cosine shaping methods were developed for the simulated nuclear signals and the actual sampled nuclear signals, and the cosine shaping, amplitude extraction, and energy spectrum construction of the digital nuclear signals were implemented in the field programmable gate array (FPGA) system. Finally, the gamma (γ) energy spectrum of 137Cs (NaI(Tl) detector) was evaluated using the different cosine shaping methods.

    Results

    The results of γ energy spectrum from 137Cs (NaI(Tl) detector) demonstrate the satisfactory performance of all three digital cosine shaping algorithms in terms of energy resolution and counting. The symmetric zero-area cosine shaping performance index is improved relative to conventional methods.

    Conclusions

    The three kinds of digital cosine shaping methods all achieve accurate cosine shaping for simulated and real nuclear signals. The three cosine shaping methods proposed in this study may be applied to shape functions in other research areas.

    Tools

    Get Citation

    Copy Citation Text

    Yong WEI, Huaiqiang ZHANG, Yunchen QIAN, He CHEN. Digital shaping methods for nuclear pulse signal based on cosine function[J]. NUCLEAR TECHNIQUES, 2023, 46(10): 100402

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Mar. 22, 2023

    Accepted: --

    Published Online: Nov. 21, 2023

    The Author Email:

    DOI:10.11889/j.0253-3219.2023.hjs.46.100402

    Topics